Overview
Olympus CLV-S400
VISERA 4K UHD xenon light source with NBI
Custom Light Source for the 4K System
· Narrow Band Imaging compatibility
· 300 W xenon lamp
· Automatic light control
Specifications
Power input | Rated voltage | 100–240 V AC; within ±10% |
Rated frequency | 50/60 Hz; within ±1 Hz |
Rated input | 500 VA |
Size | Dimensions (maximum) | 391 (W) × 162 (H) × 521 (D) mm |
Weight | 15.5 kg |
Illumination | Examination lamp | Xenon short-arc lamp (ozone-free) 300 W |
Average lamp life | Approximately 500 hours of continuous use (with intermittent use, the lamp life may vary slightly) |
Ignition method | Switching regulator |
Brightness adjustment | Light-path diaphragm control |
Cooling | Forced-air cooling |
Intensity mode | Normal or high intensity |
NBI observation | Available |
Emergency lamp | Halogen lamp (within mirror) 12 V 35 W |
Average emergency lamp life | Approximately 500 hours |
Automatic brightness control | Automatic brightness control method | Servo-diaphragm method |
Automatic exposure | 17 steps |
Indicators on front panel | Emergency lamp | Indicates absence of emergency lamp, disconnection and use of emergency lamp |
NBI | When the NBI observation mode is enabled, the observation mode indicator lights up |
Setting memory |
| Settings (except the observation mode) are stored even when the light source is OFF |
Classification (medical electrical equipment) | Type of protection against electric shock | Class I |
Degree of protection against electric shock of applied part | TYPE BF applied part |
Degree of protection against explosion | The light source should be kept away from flammable gases |